ProteOn XPR36 | Experiments
70
2. In the Autosampler Temperature area, type a temperature value in the
Temperature box.
3. Click Set.
To set the default autosampler temperature:
•
Select Options on the Tools menu and specify the autosampler
temperature in the dialog box that appears
Sensor Chips
Sensor chips are encased in a plastic cartridge for easy handling and loading
into the instrument. Each cartridge-encased chip is sealed in an aluminum
nitrogen-filled pouch to maximize shelf life. The chips should remain in the
sealed pouch until use.
The label on the sensor chip cartridge contains the following information:
•
Chip type
•
Catalog number
•
Bar code
•
Expiration date
Each sensor chip is a gold-coated glass prism containing surface chemistry
used for immobilizing biomolecules (ligands) to the chip surface. This enables
interactions with other biomolecules (analytes) to create a measurable SPR
response used in kinetic analysis and other applications. See Appendix C
Surface Plasmon Resonance for more information.
The following types of interactions can be carried out:
•
Protein-protein
•
Protein-peptide
•
Protein-small molecule
•
Protein–DNA
Summary of Contents for XPR36
Page 1: ...ProteOn XPR36 Protein Interaction Array System Protein Interaction Analysis User Manual...
Page 2: ......
Page 3: ...ProteOn XPR36 Protein Interaction Array System Version 3 1 User Manual...
Page 4: ......
Page 10: ...ProteOn XPR36 System viii...
Page 166: ...ProteOn XPR36 System Analysis 148...
Page 188: ...ProteOn XPR36 Maintenance 170...
Page 194: ...ProteOn XPR36 System ProteOn System Troubleshooting 176...
Page 226: ...ProteOn XPR36 System A ProteOn XPR36 Instrument and Peripherals 208...
Page 250: ...ProteOn XPR36 System E Ordering Information 232...
Page 264: ...ProteOn XPR36 System F Security Edition Configuration Guide 246...
Page 276: ...ProteOn XPR36 System Glossary 258...
Page 287: ......