Maintenance Wizard Instructions
165
OQ Protocol
The Maintenance screen also enables you to run ProteOn Operation
Qualification, an optional protocol that ensures the instrument is performing
within manufacturer specifications. This protocol is available from Bio-Rad.
This protocol supports the Good Practices rulings within the pharmaceutical
industry. Bio-Rad requires weekly, post-experiment, and MCM cleaning
maintenance procedures and recommends running the OQ protocol twice a
year, as shown in the table.
The OQ protocol can be run after the required quarterly maintenance to
minimize instrument downtime.
For complete instructions on running the OQ Protocol, refer to the ProteOn
XPR36 IQ/OQ Software User Manual, 100005737.
Quarterly Maintenance Schedule
Q1
Q2
Q3
Q4
Multichannel module (MCM) cleaning
X
X
X
X
Operation Qualification (OQ) Protocol
X
X
Summary of Contents for XPR36
Page 1: ...ProteOn XPR36 Protein Interaction Array System Protein Interaction Analysis User Manual...
Page 2: ......
Page 3: ...ProteOn XPR36 Protein Interaction Array System Version 3 1 User Manual...
Page 4: ......
Page 10: ...ProteOn XPR36 System viii...
Page 166: ...ProteOn XPR36 System Analysis 148...
Page 188: ...ProteOn XPR36 Maintenance 170...
Page 194: ...ProteOn XPR36 System ProteOn System Troubleshooting 176...
Page 226: ...ProteOn XPR36 System A ProteOn XPR36 Instrument and Peripherals 208...
Page 250: ...ProteOn XPR36 System E Ordering Information 232...
Page 264: ...ProteOn XPR36 System F Security Edition Configuration Guide 246...
Page 276: ...ProteOn XPR36 System Glossary 258...
Page 287: ......