ProteOn XPR36 | Experiments
86
requires at least three steps. See To edit a protocol or template: on
page 32.
9. Make EVC standards that contain cosolvent concentrations less than,
equal to, and greater than the cosolvent concentration in the analyte
samples and running buffer. Place them in the rack or microplate to be
used.
10. Start the experiment by clicking Start on the Run screen. See
Processing EV Correction Data on page 115 for information on
processing and analyzing EV corrected data.
Saving and Exporting Experiment Data
ProteOn Manager software automatically saves experiment data to the
database. For information about exporting your experiment data, see Saving
and Exporting Data on page 141.