![ProteOn XPR36 User Manual Download Page 268](http://html1.mh-extra.com/html/proteon/xpr36/xpr36_user-manual_1620131268.webp)
ProteOn XPR36 System | Glossary
250
Double Reference
Correction of analyte data using a blank injection as the reference.
Prior to applying the double reference, both the blank and analyte
data are corrected using interspot or channel referencing. Sensor
chip background is interspot reference and background contributed
by biological (experimental) sources (for example, buffer without
analyte). Subtracting both backgrounds is double referencing
Dynamic Range
Upper and lower bounds on the ability of a system (that is,
instrument, reagents, etc.) to measure. In the context of interaction
analysis, it refers to the range of SPR signals which the ProteOn
XPR36 system can measure. Dynamic range is expressed in
response units
EDAC
Also abbreviated EDC, 1-ethyl-3-(3-dimethylaminopropyl)
carbodiimide hydrochloride is used for sensor chip surface activation
(see
Activation
)
Equilibration
See
Stabilization
Equilibrium
Constant (K
D
), also known as the equilibrium dissociation constant,
characteristic of a reversible chemical reaction, measured in units of
molarity, or M. It relates the concentrations of all reactants and
products at equilibrium. It is calculated from the ratio of the
dissociation and association kinetic rate constants, such that
K
D
= k
d
/k
a
Evanescent Wave
Electromagnetic wave generated by a light wave in total internal
reflection at the interface of two media. It decays exponentially with
distance as it passes into the medium of lower refractive index
Excluded Volume
(EV) Correction
Use of high refractive index cosolvents, such as DMSO, in SPR
results in a differential bulk response between the chip's reference
surface and the interaction surface. This is due to the cosolvent
being excluded from the chip surface by the ligand and results in a
larger bulk response on the reference surface than on the interaction
surface. If not corrected, this excess response results in an error
when subtracting the reference data from the interaction data. The
excluded volume correction function calculates the excess response
from a calibration curve and then removes the excess response from
the reference data
Summary of Contents for XPR36
Page 1: ...ProteOn XPR36 Protein Interaction Array System Protein Interaction Analysis User Manual...
Page 2: ......
Page 3: ...ProteOn XPR36 Protein Interaction Array System Version 3 1 User Manual...
Page 4: ......
Page 10: ...ProteOn XPR36 System viii...
Page 166: ...ProteOn XPR36 System Analysis 148...
Page 188: ...ProteOn XPR36 Maintenance 170...
Page 194: ...ProteOn XPR36 System ProteOn System Troubleshooting 176...
Page 226: ...ProteOn XPR36 System A ProteOn XPR36 Instrument and Peripherals 208...
Page 250: ...ProteOn XPR36 System E Ordering Information 232...
Page 264: ...ProteOn XPR36 System F Security Edition Configuration Guide 246...
Page 276: ...ProteOn XPR36 System Glossary 258...
Page 287: ......