Editing Protocol Sample Details
47
To drag a sample panel:
•
Drag the selected panel to a new position. An alert asks if you want
to proceed. The dropped panel overwrites the panel in the new
position
To copy and paste a sample panel:
1. In the Samples screen, collapse all panels by clicking Collapse all
in the
Display tools group.
2. Select a panel. Copy multiple samples by selecting them while pressing
the CTRL key.
3. Select Copy
on the Edit menu.
4. Select the sample panel or panels you want to overwrite.
5. Select Paste on the Edit menu.
Note:
When copying multiple panels, you must select the same
number of panels you want to overwrite before you can paste.
Autosampler Layout
The Autosampler Layout box shows the position of samples and reagents in
the sample holder. Clicking a column of samples in the Autosampler Layout
box brings the corresponding sample panel into focus.
The sample panel in focus displays samples in different colors. Each other
sample panel is colored according to its sample panel type.
Summary of Contents for XPR36
Page 1: ...ProteOn XPR36 Protein Interaction Array System Protein Interaction Analysis User Manual...
Page 2: ......
Page 3: ...ProteOn XPR36 Protein Interaction Array System Version 3 1 User Manual...
Page 4: ......
Page 10: ...ProteOn XPR36 System viii...
Page 166: ...ProteOn XPR36 System Analysis 148...
Page 188: ...ProteOn XPR36 Maintenance 170...
Page 194: ...ProteOn XPR36 System ProteOn System Troubleshooting 176...
Page 226: ...ProteOn XPR36 System A ProteOn XPR36 Instrument and Peripherals 208...
Page 250: ...ProteOn XPR36 System E Ordering Information 232...
Page 264: ...ProteOn XPR36 System F Security Edition Configuration Guide 246...
Page 276: ...ProteOn XPR36 System Glossary 258...
Page 287: ......