Analyzing Data
131
It is possible to change the starting parameters for the fit calculation by typing
the desired values in the table. When you want to use a pre-determined value
for one of the fit parameters, change the Fit Type of this parameter to
Constant, and type the requested value in the table. For example, pre-
determined R
max
could be set to the theoretical R
max
value in low affinity
models, when there is no or limited curvature in the fitted sensorgrams. See
Immobilization Step on page 216 for theoretical R
max
calculation.
Step 3: Results Preview
When you click Next, the wizard calculates the fit. Once the fit is calculated,
the following table appears with the Report tab open.
Analyzed datasets show sensorgrams and their fitted curves in data windows.
Icons appear in the left margin of the dialog box.
•
Report
— opens the table summarizing numerical results of the
analysis
•
Residual
— opens the kinetic analysis residuals graph, which plots
the calculation residuals along the y axis and experiment time
along the x axis. Residuals are a graphical representation of the
fitting accuracy and appear only for the sensorgrams selected with
the Interaction Display Chooser during data preprocessing
Summary of Contents for XPR36
Page 1: ...ProteOn XPR36 Protein Interaction Array System Protein Interaction Analysis User Manual...
Page 2: ......
Page 3: ...ProteOn XPR36 Protein Interaction Array System Version 3 1 User Manual...
Page 4: ......
Page 10: ...ProteOn XPR36 System viii...
Page 166: ...ProteOn XPR36 System Analysis 148...
Page 188: ...ProteOn XPR36 Maintenance 170...
Page 194: ...ProteOn XPR36 System ProteOn System Troubleshooting 176...
Page 226: ...ProteOn XPR36 System A ProteOn XPR36 Instrument and Peripherals 208...
Page 250: ...ProteOn XPR36 System E Ordering Information 232...
Page 264: ...ProteOn XPR36 System F Security Edition Configuration Guide 246...
Page 276: ...ProteOn XPR36 System Glossary 258...
Page 287: ......