![ProteOn XPR36 User Manual Download Page 54](http://html1.mh-extra.com/html/proteon/xpr36/xpr36_user-manual_1620131054.webp)
ProteOn XPR36 System | Protocol Design
36
To create a protocol from rack/plate contents:
1. Select the Protocol tab, and then select the Samples screen.
2. In the Step Creation area, select Create Protocol. The Create Protocol
From Samples dialog box appears, in which you can build a cycle of
steps.
3. In the Take samples from drop-down list, select the rack or plate that is
the source of the samples.
4. If you make no other changes in this dialog box and click Create, a
cycle of the three steps, in the order of the tabs, is added to the
protocol Steps list.
To repeat an injection or block:
•
You can change the number of times to repeat an injection by
specifying the number in the Repeat Injection box in the Create
Protocol From Samples dialog box. Specifying zero in the Repeat
Injection box omits the step type from a cycle of steps
•
You can change the number of times to repeat a block by
specifying the number in the Repeat block box. A block repeats the
steps for each analyte, so you can create a cycle of steps and a
replicate of that cycle of steps
Summary of Contents for XPR36
Page 1: ...ProteOn XPR36 Protein Interaction Array System Protein Interaction Analysis User Manual...
Page 2: ......
Page 3: ...ProteOn XPR36 Protein Interaction Array System Version 3 1 User Manual...
Page 4: ......
Page 10: ...ProteOn XPR36 System viii...
Page 166: ...ProteOn XPR36 System Analysis 148...
Page 188: ...ProteOn XPR36 Maintenance 170...
Page 194: ...ProteOn XPR36 System ProteOn System Troubleshooting 176...
Page 226: ...ProteOn XPR36 System A ProteOn XPR36 Instrument and Peripherals 208...
Page 250: ...ProteOn XPR36 System E Ordering Information 232...
Page 264: ...ProteOn XPR36 System F Security Edition Configuration Guide 246...
Page 276: ...ProteOn XPR36 System Glossary 258...
Page 287: ......