![ProteOn XPR36 User Manual Download Page 239](http://html1.mh-extra.com/html/proteon/xpr36/xpr36_user-manual_1620131239.webp)
Determining the Optimum pH for Immobilization
221
Determining the Extent of Ligand
Immobilization
Changes in the SPR signal are monitored during immobilization.
Immobilization levels are determined by the signal and measured in response
units (RU), observed as the difference between the level before ligand
injection and the level after deactivation is complete.
The illustration below shows SPR response during the ligand immobilization
process. The baseline is based on the steady signal (immobilization level)
observed following deactivation.