ProteOn XPR36 System | A ProteOn XPR36 Instrument and Peripherals
200
Needle Holder with Sampling Needles
The needle arm moves the sampling needles into position for sampling and for
needle washing. The needle holder maintains the six sampling needles in the
proper configuration and is covered by a safety needle guard.
Two interchangeable needle holders with different needle spacing are
provided: one to match the spacing of microplate wells and the other to match
the spacing of sample vials in the ProteOn sample rack.
Note:
Rack/plates must match the needle holder type. If there is
a mismatch, the instrument will go into Fault state.
Changing the Needle Holder and Needles
These instructions describe how to change the sample vial and microplate
needle holder configurations.
Tools:
•
Three wrenches are provided with the ProteOn XPR36 system: 1.5
mm hex wrench, 3 mm hex wrench, and a fitting wrench
Safety Reminders
Needle tips are sharp; use caution when
handling them.
Shut down the instrument and turn off the power
before handling the autosampler needle holder
or needles. See Shutdown Procedures on
page 166.
Follow appropriate biosafety procedures.
Summary of Contents for XPR36
Page 1: ...ProteOn XPR36 Protein Interaction Array System Protein Interaction Analysis User Manual...
Page 2: ......
Page 3: ...ProteOn XPR36 Protein Interaction Array System Version 3 1 User Manual...
Page 4: ......
Page 10: ...ProteOn XPR36 System viii...
Page 166: ...ProteOn XPR36 System Analysis 148...
Page 188: ...ProteOn XPR36 Maintenance 170...
Page 194: ...ProteOn XPR36 System ProteOn System Troubleshooting 176...
Page 226: ...ProteOn XPR36 System A ProteOn XPR36 Instrument and Peripherals 208...
Page 250: ...ProteOn XPR36 System E Ordering Information 232...
Page 264: ...ProteOn XPR36 System F Security Edition Configuration Guide 246...
Page 276: ...ProteOn XPR36 System Glossary 258...
Page 287: ......