ProteOn XPR36 System | A ProteOn XPR36 Instrument and Peripherals
204
S
AMPLE
H
OLDER
S
ENSORS
Three optomechanical sample holder sensors are located at the back of the
thermal platform. The sensors detect the type of sample holder being used
and relay the information to the software.
Syringe Pumps
The instrument utilizes two syringe pumps:
•
Buffer syringe pump (right)
•
Sample syringe pump (left)
Each pump contains a set of six 500 μl glass syringes. The buffer syringe
pump mediates the flow of buffer from the buffer bottles into the multichannel
module. The sample syringe pump mediates the flow of samples from the
autosampler into the multichannel module. The sample syringe pump is also
used to flush the sample tubing during wash procedures.
Warning!
You must use the needle holder
corresponding to the type of sample holder you
are using. Failure to use the correct needle
holder will result in damage to needles and
possibly to other components.
Summary of Contents for XPR36
Page 1: ...ProteOn XPR36 Protein Interaction Array System Protein Interaction Analysis User Manual...
Page 2: ......
Page 3: ...ProteOn XPR36 Protein Interaction Array System Version 3 1 User Manual...
Page 4: ......
Page 10: ...ProteOn XPR36 System viii...
Page 166: ...ProteOn XPR36 System Analysis 148...
Page 188: ...ProteOn XPR36 Maintenance 170...
Page 194: ...ProteOn XPR36 System ProteOn System Troubleshooting 176...
Page 226: ...ProteOn XPR36 System A ProteOn XPR36 Instrument and Peripherals 208...
Page 250: ...ProteOn XPR36 System E Ordering Information 232...
Page 264: ...ProteOn XPR36 System F Security Edition Configuration Guide 246...
Page 276: ...ProteOn XPR36 System Glossary 258...
Page 287: ......