ProteOn XPR36 System | Analysis
116
3. (Optional) To exclude data points, maximize the calibration plot with the
erroneous data point by double-clicking it.
Note:
At least three data points (solid) must be selected from
each calibration plot.
4. On the plot, click the data points you want to exclude. The excluded
data point is represented by an empty circle. Note the resulting change
in fit in the R
2
table.
5. In the Choose calibration steps list, select a minimum of three excluded
volume correction (EVC) calibration steps if they are not already
selected) and click Apply.
6. Click Finish to subtract the corrected reference data from the selected
step data.
If you change the channel reference after finishing the EVC, the following
dialog box appears:
•
Use New Reference With No Correction
— Select this option to
switch from an excluded volume corrected channel reference to a
non-excluded volume corrected channel reference
•
Calibrate a New Correction For This Reference
— Select this
option to switch from an excluded volume corrected channel
Summary of Contents for XPR36
Page 1: ...ProteOn XPR36 Protein Interaction Array System Protein Interaction Analysis User Manual...
Page 2: ......
Page 3: ...ProteOn XPR36 Protein Interaction Array System Version 3 1 User Manual...
Page 4: ......
Page 10: ...ProteOn XPR36 System viii...
Page 166: ...ProteOn XPR36 System Analysis 148...
Page 188: ...ProteOn XPR36 Maintenance 170...
Page 194: ...ProteOn XPR36 System ProteOn System Troubleshooting 176...
Page 226: ...ProteOn XPR36 System A ProteOn XPR36 Instrument and Peripherals 208...
Page 250: ...ProteOn XPR36 System E Ordering Information 232...
Page 264: ...ProteOn XPR36 System F Security Edition Configuration Guide 246...
Page 276: ...ProteOn XPR36 System Glossary 258...
Page 287: ......