EMC Guidance and Manufacturer’s Declaration 17-1
17
EMC Guidance and
Manufacturer’s Declaration
The system complies with the EMC standard IEC 60601-1-2: 2007.
WARNING:
1. The use of unapproved accessories may diminish system
performance.
2. Use of components, accessories, probes, and cables other than
those specified may result in increased emission or decreased
immunity of system.
3. The system or its components should not be used adjacent to
or stacked with other equipment. If adjacent or stacked use is
necessary, the system or its components should be observed
to verify normal operation in the configuration in which it will
be used.
4. Operation of system, in the case that the patient physiological
signal is lower than the minimum amplitude or value specified
in the product specifications, results may be inaccurate (results
can be obtained when the HR is in the range of 30-250 bmp or
when the QRS wave amplitude is between 0.5-5 mV.)
NOTE:
1.
The system needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information provided
below.
2.
Other devices may interfere with this system even though they meet the
requirements of
CISPR
in the corresponding standards.
3.
Preventing conducted RF immunity. Due to technological limitations, the
conducted RF immunity level are limited to 3 Vrms level, conducted RF
interference above 3Vrms may cause wrong diagnosis and measurements.
We suggest that you position system further from sources of conducted RF
noise.
4.
Portable and mobile RF communications equipment can affects system.
See tables 1, 2, 3, and 4 below.
If the system is operated within the electromagnetic environment listed in Table 2 and
Table 3, the system will remain safe and will provide the following basic performances:
Imaging;
Doppler acoustic spectral displaying;
Taking measurements;
Patient information;
Date/time information.
Summary of Contents for M5 Exp
Page 2: ......
Page 12: ......
Page 41: ...System Overview 2 11 UMT 200 UMT 300...
Page 246: ...12 2 Probes and Biopsy V10 4B s CW5s 4CD4s P12 4s 7L4s L12 4s P7 3s L14 6Ns P4 2s CW2s...
Page 286: ......
Page 288: ......
Page 336: ......
Page 338: ......
Page 357: ...P N 046 008768 00 V1 0...