12-30 Probes and Biopsy
NGB-016
1. Put on the probe sheath.
2. Select a proper needle-guided bracket, and match the groove with the tab of the
probe. Mount the bracket onto the probe.
3. Screw the pinch nut of the needle-guided bracket to confirm that the needle-guided
bracket is properly installed on the probe.
4. Select the appropriate guiding block, and fit it into the groove on the angle block.
5. Screw the nut on the guiding block to secure the guiding block with the needle-guided
bracket.
6. Insert the needle of the same specification with that of the guiding block into the
guiding hole.
NGB-018
(1) Put the sterile probe sheath on.
(2) Select a suitable needle-guided bracket and match the groove to the tab of the
transducer. Mount the bracket onto the transducer. The needle-guided brackets
may be different from each other, but the methods are the same.