5-82 Image Optimization
Item
Function Description
Protocol name
Enter the protocol name.
Stage
Set stages for the protocol.
Trigger
Set the trigger type.
WMS model
Set the methods of chamber segments division.
Loop usage
Displays the acquired loop number as well as the total usable loop
number.
View
Set the views for the each stage.
Standard views
Set the standard view.
Import
Import a protocol.
Export
Export a protocol.
New
Create a new protocol.
Load
Load an existing protocol.
Save
Save the change (s) for the protocol.
Delete
Delete the protocol.
Quit
To remove the Protocol Editor dialogue box.
Creating a Stress Echo Protocol
To create a stress echo protocol:
1. Click [New] button on the right of the Protocol Editor dialog box.
2. Enter the protocol name in the Protocol Name box at the top.
3. For each phase in the protocol:
Select “new” in the Stage list.
Enter a phase name in the Description box.
Select the required option from the Clip Capture drop-down list.
Select the number of loops to acquire (per view in the selected phase) in the Loops
list (for non-continuous stages).
Select the type: exercise or drug.
4. For each view (all views display for each phase):
Select “new” in the View list.
Select a standard view from the Standard View list.
5. Click [Save] to save changes and then click [Quit] to remove the Protocol Editor
dialog box from the screen.
Editing or Deleting a Stress Echo Protocol
To delete a protocol:
Select the protocol or one of its component phases or views and then click [Delete].
To Edit a Stress Echo Protocol:
You cannot edit a protocol during Acquire Mode.
Method one: