EMC Guidance and Manufacturer’s Declaration 17-3
TABLE 2
GUIDANCE AND MINDRAY DECLARATION—ELECTROMAGNETIC IMMUNITY
The system is intended for use in the electromagnetic environment specified below. The
customer or the user of system should assure that it is used in such an environment.
IMMUNITY TEST
IEC 60601
TEST LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC
ENVIRONMENT-GUIDANCE
Electrostatic
Discharge(ESD)
IEC 61000-4-2
±6 kV contact;
±8 kV air
±6 kV contact;
±8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be at
least 30%.
Electrical fast
Transient / burst
IEC 61000-4-4
±2 kV for power
supply lines;
±2 kV for power
supply lines;
Mains power quality should
be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV line(s) to
line(s);
±2 kV line(s) to
earth
±1 kV line(s) to
line(s);
±2 kV line(s) to
earth
Mains power quality should
be that of a typical
commercial or hospital
environment.
Voltage dips,
Short interruptions
and voltage
variation on power
supply input
voltage
IEC 61000-4-11
<5% U
T
(>95%
dip in U
T
) for
0.5 cycle
40% U
T
(60%
dip in U
T
) for 5
cycle
70% U
T
(30%
dip in U
T
) for 25
cycle
<5% U
T
(>95%
dip in U
T
) for 5
sec
<5% U
T
(>95%
dip in U
T
) for 0.5
cycle
40% U
T
(60% dip
in U
T
) for 5 cycle
70% U
T
(30% dip
in U
T
) for 25 cycle
<5% U
T
(>95%
dip in U
T
) for 5
sec
Mains power quality should
be that of a typical
commercial or hospital
environment. If you require
continued operation during
power mains interruptions, it
is recommended that our
product be powered from an
uninterruptible power supply
or a battery.
Power frequency
(50/60 HZ)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE:
U
T
is the A.C. mains voltage prior to application of the test level.
Summary of Contents for M5 Exp
Page 2: ......
Page 12: ......
Page 41: ...System Overview 2 11 UMT 200 UMT 300...
Page 246: ...12 2 Probes and Biopsy V10 4B s CW5s 4CD4s P12 4s 7L4s L12 4s P7 3s L14 6Ns P4 2s CW2s...
Page 286: ......
Page 288: ......
Page 336: ......
Page 338: ......
Page 357: ...P N 046 008768 00 V1 0...