DICOM 11-15
DICOM query/retrieve setting items are described as follows:
Name
Description
Device
After you set the server (s) in DICOM Server Preset screen, the name (s) will
appear in the drop-down list, select the name of the server.
Service
Name
Default: xxx-Store, and it can be modified
AE Title
Application Entity title
Port
DICOM communication port, 104 is default. Here, the port should be
consistent with that of the query/retrieve server port.
Maximum
Retries
Reserved feature.
Interval
Time(s)
Reserved feature.
Timeout
Value: 5
-
60s, in increments of 5s, and 15 by default.
Add
Add the DICOM service to the service list
Cancel
Click to cancel parameter setting
Delete
Click to delete the selected service in the service list
Update
Select an item in the service list, change the parameters in the above area,
and click [Update] to update the item in the service list.
Verify
Click to verify if the two DICOM application entities are normally connected.
Default
Select an item in the service list, click [Default] and you can see “Y” in the
Default column.
Exit
Click to exit the screen.