12-32 Probes and Biopsy
NGB-027
1. Put on the sterile transducer sheath.
2. Open the clamp. Insert the front clamp to the front groove.
3. Push the biopsy forward (arrow’s direction) until the locating pole inserting into the
location hole. Turn the lower clamp against the intra-cavity probe. Tighten the nut to
lock the biopsy (arrow’s direction).
CAUTION:
Ensure that all guide parts are seated properly prior to
performing a biopsy.
12.2.5 Biopsy Menu
Move the cursor onto the image menu; navigate the cursor to [Other] item and press
<Set> or the multifunctional knob. Click [Biopsy] to enter the biopsy menu.
Select biopsy bracket angle
If the needle-guided bracket supports more than one biopsy angle, you can select the
angle from the drop-down list (the last two numerals).