1-10 Safety Precautions
NOTE:
1. Read the following precautions to prevent the probe from malfunction:
Clean and disinfect the probe before and after each examination.
After the examination, wipe off the ultrasound gel thoroughly.
Otherwise, the ultrasound gel may solidify and the image quality
would be degraded.
2. Ambient conditions:
To prevent the probe from being damaged, do not use it where it will be
exposed to:
Direct sunlight or X-rays
Sudden changes in temperature
Dust
Excessive vibration
Heat generators
Use the probes under the following ambient conditions:
ambient temperature: 0
℃~
40
℃
relative humidity:30%
~
85% (no condensation)
atmospheric pressure:700 hPa
~
1060 hPa.
Use the L16-4Hs and D7-2s probe under the following ambient conditions:
ambient temperature: 10°C
~
40°C
relative humidity: 30%
~
85% (no condensation)
atmospheric pressure: 700 hPa
~
1060 hPa
Use the DE11-3s probe under the following ambient conditions:
ambient temperature: 18°C
~
30°C
relative humidity: 20%
~
85% (no condensation)
atmospheric pressure: 700 hPa
~
1060 hPa
Use the C6-2Gs probe under the following ambient conditions:
ambient temperature: 0°C
~
40°C
relative humidity: 20%
~
85% (no condensation)
atmospheric pressure: 700 hPa
~
1060 hPa
3. Repeated disinfection will eventually damage the probe, please check the
probe's performance periodically.