3-6 System Preparation
3.4
Connecting /Disconnecting a Probe
CAUTION:
1.
When connecting or disconnecting a probe, place it in a
proper position, to prevent the probe from falling off or
becoming damaged.
2. Only use the probes provided by Mindray. Aftermarket
probes may result in damage or cause a fire.
3.4.1
Connecting a Probe
WARNING:
The probes, cables and connectors are in proper operating
order and free from surface defects, cracks and peeling.
Using a defective probe may cause electric shock.
1. Keep the cable end of the probe to the right side of the system, and insert the
connector into the port of the system, and then press in fully. See the figure below.
2. Toggle the locking lever to the top position.
3. Place the cable properly to avoid being treaded or wrapping with other devices. DO
NOT allow the probe head to hang free.
3.4.2
Disconnecting a Probe
1. Toggle the locking lever to the bottom position to unlock the connector of the probe.
2. Pull the probe connector straight out as shown in the figure below.
Locking lever