Measurement 8-5
8.4
Measurement Accuracy
Table 1 Error of 2D Images
Parameter
Value Range
Error
Distance
Full screen
Within ±3%; or when the measured value is
less than 40 mm, the error is less than1.5 mm.
Area (Trace)
Full screen
Within ±7%; or when the measured value is
less than 16 cm
2
, the error is less than 1.2 cm
2
.
Area (ellipse,
circle)
Full screen
Within ±7%; or when the measured value is
less than 16 cm
2
, the error is less than 1.2 cm
2
.
Angle
Full screen
Within ±3%.
Table 2 Volume Measurements
Parameter
Value
Range
Error
Volume
Full screen
Within ±10 %; or when the measured value is less than 64 cm
3
,
the error is less than 6.4 cm
3
.
Table 3 Time/Motion Measurements
Parameter
Value Range
Error
Distance
Full screen
Within ±3 %; or when the measured value is
less than 40mm, the error is less than 1.5mm.
Time
Timeline Display
Within ±2%.
Heart rate
Timeline Display
Within ±4%.
Table 4 Doppler Velocity Measurements
Parameter
Value Range
Error
Velocity (PW mode)
Maximum ≥2.5m/s
When angle≤60º, ≤5%.
Velocity (CW mode)
Maximum ≥2.5m/s
When angle≤60º, ≤5%.
The above measurements are not applicable for the pencil probe (CW2s and CW5s).