1-6 Safety Precautions
5.
Do not connect this system to outlets with the same
circuit breakers and fuses that control the current of
devices such as life-support systems. If this system
malfunctions and generates over current, or when there
is an instantaneous current at power ON, the circuit
breakers and fuses of the building’s supply circuit may
be tripped.
6.
Always keep the system dry. Avoid transporting this
system quickly from a cold place to a warm place;
otherwise condensation or water droplets may form
allowing a short circuit and possible electric shock.
7.
If the circuit protector is tripped, it indicates that the
system or a peripheral device was improperly shut down
and the system is unstable. You cannot repair the
system under this circumstance and must call the
Mindray Customer Service Department or sales
representative.
8.
There is no risk of high-temperature burns during
normal ultrasound examinations. It is possible for the
surface temperature of the probe to exceed the body
temperature of a patient due to environmental
temperature and exam mode combinations. Apply the
probe only for a period of time required for the purpose
of diagnosis.
9.
The system and its accessories are not disinfected or
sterilized prior to delivery. The operator is responsible
for the cleaning and disinfection of probes and
sterilization of biopsy brackets according to the
manuals, prior to the use.
All items must be thoroughly processed to completely
remove harmful residual chemicals, which will not only
harmful to the human body, but also damage the
accessory.
10. It is necessary to press [End Exam] to end the current
scan that is in progress and clear the current Patient
Information field. Otherwise, new patient data may be
combined with the previous patient data.
11.
DO NOT connect or disconnect the system’s power cord
or its accessories (e.g., a printer or a recorder) without
turning OFF the power first. This may damage the
system and its accessories or cause electric shock.
12. If the system is powered off improperly during
operation, it may result in data damage of the system’s
hard disk or system failure.
13. Do not use the system to examine a fetus for a long
period of time.
14. Do not use a USB memory device (e.g., a USB flash
drive, removable hard disk) which has unsafe data.
Otherwise, system damage may result.