12-12 Probes and Biopsy
For details about additional infection control procedures or disinfection information,
contact Mindray North America Customer Service [Tel: (1-800) 288 2121, (1-201) 995
8000 ].
Sterilization
For intra-operative probes, they have to be sterilized after completing each examination.
1. Wear sterile gloves to prevent infection.
2. Clean the probe before sterilizing it. MINDRAY recommends the following solutions to
sterilize the probe.
Hydrogen Peroxide and Peroxyacetic Acid -based sterilization solution
Trade Name
Chemical Name
Procedures
Minncare
®
Cold
Sterilant
22% Hydrogen
Peroxide
4.5% Peroxyacetic
Acid
Dilute the sterilant with sterilized purified
water (1:20).
Immersed time: 11 hours.
Temperature: 20
℃
-25
℃
.
Please refer to the instructions provided by
the manufacturer of the solution for details.
Before safety and performance is affected, probe 7LT4s/L16-4Hs can be sterilized by
Minncare COLD STERILANT for at least 135 times (11 hours for one time).
Glutaraldehyde-based sterilization solution
Trade Name
Chemical Name
Procedures
Cidex Activated
Glutaraldehyde
Solution
2.2-2.7%
Glutaraldehyde
Soak the transducer into the activated
solution for 10 hours (
20-25
℃
)
Please refer to the instructions provided by
the manufacturer of the solution for details.
Before safety and performance is affected, probe 7LT4s/L16-4Hs can be sterilized by
Cidex Activated
Glutaraldehyde
Solution for at least 217 times (10 hours for one time).
Refer to the instructions provided by the chemical manufacturer concerning
concentration of the sterilization solution, method of sterilization and dilution and
cautions during use.
Do not soak the probe connector or the cable near it into water or any solution.
Follow local regulations when selecting and using the sterilization solution.
3. Rinse the probe with plenty of sterile water (about 2 gallons) for at least 1 minute to
remove all chemical residues on it. Or, follow the rinsing method recommended by the
sterilization solution manufacturer to rinse the probe.
4. Wipe off the water on the probe with sterile cloth or gauze after rinsing it. Do not dry
the probe by heating.
Immerse the probe in the solution (take 7LT4s as an illustration)
Summary of Contents for M5 Exp
Page 2: ......
Page 12: ......
Page 41: ...System Overview 2 11 UMT 200 UMT 300...
Page 246: ...12 2 Probes and Biopsy V10 4B s CW5s 4CD4s P12 4s 7L4s L12 4s P7 3s L14 6Ns P4 2s CW2s...
Page 286: ......
Page 288: ......
Page 336: ......
Page 338: ......
Page 357: ...P N 046 008768 00 V1 0...