5-20 Image Optimization
Operation
Select the value through the [Priority] item in the soft menu or menu.
The adjusting range is 0-100%.
Effects
The higher the value, color signals are prior to be displayed; while the
lower the value, grayscale signals are prior to be displayed.
iTouch
Description
To optimize image parameters as per the current tissue characteristics for
a better image effect.
Operation
Press <iTouch> on the control panel to turn on the function, the symbol of
which will be displayed in the image parameter area of the screen.
HR Flow (High Resolution Flow)
Description
This function enhances the micro vessel imaging effect, and can be used to
analyze the tissue blood supply condition.
Operations
Set [HR Flow] on the menu in the top-left corner of the screen to “On” to
turn on HR Flow status.
The parameters in HR Flow mode are independent from those in Color
mode.
To use HR Flow feature, you need to select probe L12-4s/ 7L4s in Thyroid
exam mode or C5-2s/ 3C5s in Fetal Cardiac exam mode.
Smart Track
Description
To optimize image parameters as per the current tissue characteristics for
a better image effect. The angle and the position of the ROI are adjusted
after the function is enabled. The area is tracked without being affected by
the dynamic moves.
Operation
Tap [Smart Track] on the soft menu under Color/Power mode. The vessels
lay in the middle of the ROI.
Impacts
The probe L12-4s and 7L4s for Vascular, carotid and EM VAS exam modes
supports the smart track.