System Preparation 3-5
3.3.3
Standby
Definition of standby: in a normal environmental temperature and humidity condition, the
system is connected with only one probe while no other external devices or extend
modules are connected, only 3V3_STB and 5V_STB power supply is available and in the
meantime, the standby indicator is in orange.
When the battery capacity is charged to the full capacity, the standby time of the system is
no less than 64 hours.
To enter standby:
Fully fold the LCD display and wait for 30 sec, then the system enters the standby
status.
Open [Setup]->[System Preset]->“General” to set the time for screensaver and
standby. If there is no operation, the system enters screensaver after the waiting
time set for screensaver, and if there is still no operation, the system enters
standby after the waiting time for standby.
Press the power button and select “Standby”.
When the system enters the standby status, the power button backlight turns orange.
To exit standby
Unfold the LCD display.
Press the power button.
When the system enters the standby status, if need to power off:
Long press the power button for several seconds.
Press the power button to exit the standby status and then power off the system.
NOTE:
1. Power off the system if you will not use the system for a long period of time
(including storage/ transportation condition), and you should not allow the
system in standby status, otherwise the batteries will be out of power and
permanently damaged.
2. If you will not use the system for a long period of time, DO NOT leave the
system in the standby status, you should shut down the system, disconnect
power adapter, mains power, and turn off powers of all connected
peripherals.