16 Probes and Biopsy
Operator’s Manual
16 - 25
Perform the following procedure:
1.
Install the needle-guided bracket:
a.
Put on the sterile probe sheath.
b.
Hold the probe by one hand, select the proper needle-guided bracket, and hold it with the
other hand. Match the groove of the bracket with the tab of the probe. Amount the bracket
onto the probe.
c.
Screw the pinch nut of the needle-guided bracket to confirm that the needle-guided
bracket is properly installed on the probe.
2.
Adjust the needle angle to the proper shift as required:
a.
Loosen the angle pinch nut.
b.
Adjust the angle block to the desired level.
c.
Tighten the angle pinch nut.
3.
Insert the biopsy needle:
a.
Adjust the dial scale to the required needle type shift, and then screw the needle fixing nut
to lock the dial scale. (To adjust the dial scale you have to loosen the needle fixing nut
first.)
b.
Pull the lock pin and close the V-shaped cover to fix the lock pin in the groove of the
needle type adjusting base, so as to install the needle into the guiding hole.
4.
Release the needle from the bracket:
5
Needle fixing nut
6
V-shaped cover
7
Angle adjusting base
8
Angle shift sign
9
Angle pinch nut
10
Angle block
11
Pinch nut
Summary of Contents for Anesus ME7T
Page 2: ......
Page 58: ...This page intentionally left blank ...
Page 154: ...This page intentionally left blank ...
Page 164: ...This page intentionally left blank ...
Page 182: ...This page intentionally left blank ...
Page 190: ...This page intentionally left blank ...
Page 208: ...This page intentionally left blank ...
Page 254: ...This page intentionally left blank ...
Page 264: ...This page intentionally left blank ...
Page 280: ...This page intentionally left blank ...
Page 311: ......
Page 312: ...P N 046 018839 00 5 0 ...