6 - 24
Operator’s Manual
6 Image Acquisition
2.
Move the probe to capture an appropriate image of the long axis view of the left ventricular
near the sternum.
3.
Press the <Measure> key to enter the application measurement. Select the “LVOT Diam” item
and the measurement cursor is displayed on the main screen.
a.
Move the cursor to the starting point and press <Set> to confirm.
b.
Move the cursor to the end point.
Press <Clear> to cancel setting the starting point for measurement and <Update> to
switch the active end and fixed end of the caliper.
c.
Press <Set> to set the end point.
For more information about measurement, see
Advanced Volume
.
4.
Press the <Freeze> key to unfreeze the image. Move the probe to capture an appropriate image
of the apical five chamber view.
5.
Tap [Smart VTI] on the touch screen or the user-defined key to enter the Smart VTI mode.
The system will:
–
Automatically trace the color VOI, and PW sampling line and sampling volume.
–
Automatically recognize the cardiac cycle: when there are ECG inputs, the ECG signals
are preferred; when there are no ECG signals, the system automatically starts calculation.
–
Trace the LVOT spectrum in a cardiac cycle in real time to gain the VTI, HR, and CO
results of the LVOT.
6.
If necessary, adjust the color VOI, and PW sampling line and sampling volume:
–
Use the trackball/trackpad and <Set> keys to adjust the color VOI position and size, and
the PW sampling line position.
–
Tap the PW tab to enter the PW mode, and adjust the PW sampling volume.
After adjustment, the LVOT spectrum tracing and calculation results display in real time.
7.
Press the <Freeze> key to freeze the image, and check the Smart VTI calculation results.
If necessary, you can edit the LOVT spectrum manually:
a.
Tap [Edit VTI] on the touch screen.
b.
Press the <Cursor> key to activate the cursor.
c.
Move the cursor to the spectrum tracing line and press the <Set> key.
d.
Use the trackball/trackpad to trace a line.
After tracing, the calculation results display in real time.
8.
Tap [Save VTI] to save the Smart VTI calculation results of the current cardiac cycle to the
report.
9.
Tap [Graph] to check the quantitative index trend curve.
VTI, SV, and CO have different trend curves and are differentiated by colors. A maximum of 6
sets of Smart VTI results can be saved and displayed.
Summary of Contents for Anesus ME7T
Page 2: ......
Page 58: ...This page intentionally left blank ...
Page 154: ...This page intentionally left blank ...
Page 164: ...This page intentionally left blank ...
Page 182: ...This page intentionally left blank ...
Page 190: ...This page intentionally left blank ...
Page 208: ...This page intentionally left blank ...
Page 254: ...This page intentionally left blank ...
Page 264: ...This page intentionally left blank ...
Page 280: ...This page intentionally left blank ...
Page 311: ......
Page 312: ...P N 046 018839 00 5 0 ...