18 System Maintenance
Operator’s Manual
18 - 5
•
If you spill liquid on the main unit, disconnect the power supply, dry the
main unit
,
and contact your service personnel.
•
Never use abrasive materials (such as steel wool or silver polish), or
erosive cleaners (such as acetone or acetone-based cleaners).
•
Dilute and use the disinfectants according to the manufacturer's
instructions.
•
Check the system after disinfection. If there is any sign of damage, remove
it from use.
•
Follow local regulations when selecting and using the disinfectant.
•
Disinfectants listed in this section are used for disinfecting the housing of
the main unit and the monitor only, not for disinfecting the probes or the
trolley.
•
During cleaning, wear medical gloves to prevent infection.
Compatible Disinfectants
The following table lists compatible disinfectants.
Disinfection Procedures
NOTE:
The system has passed puncture test and leakage current test, and thus can be disinfected on power-
on status: tap > bar in the left side of the touch screen to enter the Exam Main Menu screen and then
select
to lock the system for 10 seconds to disinfect the system while the power is “On”.
After cleaning, disinfect the system as follows.
1.
Wear medical gloves to prevent infection.
2.
Disinfect the main unit with disinfectant wipe or solution.
Follow the disinfectant manufacturer’s recommended contact time and method.
Only the areas highlighted in gray color as shown in the following figure can be disinfected.
Do not disinfect the top cover, bottom cover, side panels around the main unit, manufacturer
logo, any visible sockets or interfaces (such as probe socket, ventilation holes, dust-proof
cover, loudspeaker, sockets or interfaces in the IO panel and power supply panel).
Product Name
Manufacturer
Type
75% alcohol
/
Solution
99.5% isopropyl alcohol
/
Solution
SONO
TM
ULTRASOUND
WIPES
Advanced Ultrasound
Solutions Inc.
Wipe
Schulke mikrozid
®
Sensitive
Wipes
Schulke
Wipe
Mikrobac Tissues
Bode Chemie Gmbh
Wipe
Sani-Cloth
®
Plus
PDI
Wipe
Summary of Contents for Anesus ME7T
Page 2: ......
Page 58: ...This page intentionally left blank ...
Page 154: ...This page intentionally left blank ...
Page 164: ...This page intentionally left blank ...
Page 182: ...This page intentionally left blank ...
Page 190: ...This page intentionally left blank ...
Page 208: ...This page intentionally left blank ...
Page 254: ...This page intentionally left blank ...
Page 264: ...This page intentionally left blank ...
Page 280: ...This page intentionally left blank ...
Page 311: ......
Page 312: ...P N 046 018839 00 5 0 ...