4 - 40
Operator’s Manual
4 Setup
If you require other maintenance functions, please contact Mindray Customer Service Department
or sales representative.
4.10.1 Option
The system enters the Option page after entering the Maintenance screen. In the Option list, the
system lists all the system-supported options and their installation status (not installed or installed).
To install and uninstall:
•
Click [Install] to begin installing a disabled option.
•
Click [Uninstall] to begin uninstalling a previously-installed option.
Please contact the Mindray Customer Service Department or a sales representative for details.
4.10.2 Exporting Setup Data
This function is used to write all setup data of the system into a disk for backup.
Perform the following procedure:
1.
Select the target module.
2.
Click [Export] to open the [Export Data] screen.
3.
Select the path to save the data.
4.
Select the exported file and type as PDP and click [OK].
4.10.3 Importing Setup Data
This function is used to import the existing setup data to the setup data memory of the system. The
system will reset and operate according to the setup preferences that were imported.
Perform the following procedure:
1.
Click [Import] to open the Load Data screen.
2.
Select the imported file.
3.
Click [OK], a progress bar will appear and the setup data is imported to the specified path.
4.
To restore the factory setup data, click [Load Factory] on the right side of the screen.
4.10.4 Probe Check
This function is used to check the condition of transducer elements to evaluate the probe
performance.
Before performing the probe check, ensure the probe head must be clean and the probe is in non-
scanning state.
Click [Probe Check] to enter the Probe Check screen:
•
If a transducer element is in malfunction, it is displayed as a red spot.
•
If a transducer element functions well, it is displayed as a green spot.
You can export the result image to the external device.
4.10.5 Other Settings
Other preset settings are described as follows:
Item
Description
Load Factory
Local factory default settings.
Summary of Contents for Anesus ME7T
Page 2: ......
Page 58: ...This page intentionally left blank ...
Page 154: ...This page intentionally left blank ...
Page 164: ...This page intentionally left blank ...
Page 182: ...This page intentionally left blank ...
Page 190: ...This page intentionally left blank ...
Page 208: ...This page intentionally left blank ...
Page 254: ...This page intentionally left blank ...
Page 264: ...This page intentionally left blank ...
Page 280: ...This page intentionally left blank ...
Page 311: ......
Page 312: ...P N 046 018839 00 5 0 ...