Operator’s Manual
I - 1
I
EMC Guidance and
Manufacturer’s Declaration
The system complies with the EMC standard IEC 60601-1-2: 2014.
Intended Environments: HOME HEALTHCARE ENVIRONMENT (except for near active HF
SURGICAL EQUIPMENT and the RF shielded room of an ME SYSTEM for magnetic resonance
imaging).
WARNING
•
The use of unapproved accessories may diminish system performance.
•
Use of components, accessories, probes, and cables other than those
specified may result in increased emission or decreased immunity of
system.
•
Operation of system, in the case that the patient physiological signal is
lower than the minimum amplitude or value specified in the product
specifications, results may not be obtained (results can be obtained when
the HR is in the range of 30-250 bmp or when the QRS wave amplitude is
between 0.5-5 mV.)
•
Use of this equipment adjacent to or stacked with other equipment should
be avoided because it could result in improper operation. If such use is
necessary, this equipment and the other equipment should be observed to
verify that they are operating normally.
•
Use of accessories, transducers and cables other than those specified or
provided by the manufacturer of this equipment could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
•
Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than 30
cm (12 inches) to any part of the system, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this equipment
could result.
•
Keep a distance of at least 20 cm away from the monitor when WiFi
function is in use.
NOTE:
This equipment has been tested and found to comply with the limits for a Class B digital device,
pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection
against harmful interference in a residential installation. This equipment generates, uses and can
radiate radio frequency energy and, if not installed and used in accordance with the instructions,
may cause harmful interference to radio communications. However, there is no guarantee that
Summary of Contents for Anesus ME7T
Page 2: ......
Page 58: ...This page intentionally left blank ...
Page 154: ...This page intentionally left blank ...
Page 164: ...This page intentionally left blank ...
Page 182: ...This page intentionally left blank ...
Page 190: ...This page intentionally left blank ...
Page 208: ...This page intentionally left blank ...
Page 254: ...This page intentionally left blank ...
Page 264: ...This page intentionally left blank ...
Page 280: ...This page intentionally left blank ...
Page 311: ......
Page 312: ...P N 046 018839 00 5 0 ...