C Electrical Safety Inspection
Operator’s Manual
C - 3
•
normal polarity with open neutral (Single Fault Condition);
•
reverse polarity with open neutral (Single Fault Condition).
LIMITS
•
For ES 60601-1:
–
300
μ
A in Normal Condition.
–
1000
μ
A in Single Fault Condition.
•
For IEC 60601-1:
–
500
μ
A in Normal Condition.
–
1000
μ
A in Single Fault Condition.
C.6
Enclosure Leakage Test
The following outlet conditions apply when performing the Enclosure Leakage test.
•
normal polarity (Normal Condition);
•
reverse polarity (Normal Condition);
•
normal polarity with open neutral (Single Fault Condition);
•
reverse polarity with open neutral (Single Fault Condition);
•
normal polarity with open earth (Single Fault Condition);
•
reverse polarity with open earth (Single Fault Condition).
LIMITS
•
For ES 60601-1
–
100
μ
A in Normal Condition.
–
300
μ
A in Single Fault Condition.
•
For IEC 60601-1:
–
100
μ
A in Normal Condition.
–
500
μ
A in Single Fault Condition.
C.7
Patient Leakage Current
Patient leakage currents are measured between a selected applied part and mains earth. All
measurements have a true RMS only.
The following outlet conditions apply when performing the Patient Leakage Current test.
•
normal polarity (Normal Condition);
•
reverse polarity (Normal Condition);
•
normal polarity with open neutral (Single Fault Condition);
•
reverse polarity with open neutral (Single Fault Condition).
•
normal polarity with open earth (Single Fault Condition);
•
reverse polarity with open earth (Single Fault Condition).
LIMITS
For BF
applied parts:
•
100
μ
A in Normal Condition.
•
500
μ
A in Single Fault Condition.
Summary of Contents for Anesus ME7T
Page 2: ......
Page 58: ...This page intentionally left blank ...
Page 154: ...This page intentionally left blank ...
Page 164: ...This page intentionally left blank ...
Page 182: ...This page intentionally left blank ...
Page 190: ...This page intentionally left blank ...
Page 208: ...This page intentionally left blank ...
Page 254: ...This page intentionally left blank ...
Page 264: ...This page intentionally left blank ...
Page 280: ...This page intentionally left blank ...
Page 311: ......
Page 312: ...P N 046 018839 00 5 0 ...