2 System Overview
Operator’s Manual
2 - 9
2.8
U-Bank Overview
WARNING
•
Only technical professionals from Mindray or engineers authorized by
Mindray after training can perform batteries in the U-Bank installation and
uninstallation.
•
If you need to change the batteries in the U-Bank, please contact your
sales representative.
Specifications are as follows:
•
Voltage: 19V DC
•
Capacity: 26400 mAh or 13200 mAh
Under power off or standby status, charging time of the battery from capacity 0 to 100% takes
about 5 hours.
U-Bank environmental conditions are as follows:
•
Operating conditions
–
Ambient temperature: 0 °C ~ 35 °C
–
Relative humidity: 20% ~ 85% (no condensation)
–
Atmospheric pressure: 700 hPa ~ 1060 hPa
•
Storage and transportation conditions
–
Ambient temperature: -20 °C ~ 55 °C
–
Relative humidity: 20% ~ 95% (no condensation)
–
Atmospheric pressure: 700 hPa ~ 1060 hPa
10.
Control Panel
Operator-system interface or control.
NOTE:
The buttons and knobs on the control panel may be different for
different product models. Please refer to the control panel of the
product for the actual buttons and knobs.
11.
Probe port
Connects a probe to the main unit; or connects a probe extend
module.
12.
Probe locking lever Locks or unlocks the probe connected with the main unit:
•
: locked symbol.
•
: unlocked symbol.
13.
Power input port
Connects the power adapter.
No.
Name
Description
Summary of Contents for Anesus ME7T
Page 2: ......
Page 58: ...This page intentionally left blank ...
Page 154: ...This page intentionally left blank ...
Page 164: ...This page intentionally left blank ...
Page 182: ...This page intentionally left blank ...
Page 190: ...This page intentionally left blank ...
Page 208: ...This page intentionally left blank ...
Page 254: ...This page intentionally left blank ...
Page 264: ...This page intentionally left blank ...
Page 280: ...This page intentionally left blank ...
Page 311: ......
Page 312: ...P N 046 018839 00 5 0 ...