16 - 40
Operator’s Manual
16 Probes and Biopsy
•
The magnetizer should be used with the needle cap. Disinfect the
magnetizer after use.
•
The needle cap is a sterile, single use device. To maintain procedural
sterility, it is important that the needle cap is isolated from non-sterile
objects.
CAUTION
•
Keep the magnetizer away from any device that is sensitive to magnetic
fields. Never place or hold the magnetizer within 1 m of the probe.
•
The eSpacial Navi system supports only the needle types listed on the
needle list. For details, see “16.4.1 Interface”.
•
Keep the L11-3VNs away from the source of disturbance, such as metal
objects and magnetized objects
.
Do not use or store L11-3VNs in the
vicinity of strong electric or magnetic fields
.
NOTE:
•
When used under optimal operating conditions by an experienced medical professional, the
eSpacial Navi system can achieve precision as high as ±2.4 mm.
•
Place the magnetizer in the proper position to prevent the magnetizer from falling off or
becoming damaged.
•
DO NOT make the magnetizer overheated (more than 50 °C) during cleaning and
disinfections. High temperature may cause the magnetizer to become deformed or damaged.
16.4.1 Interface
In Plane Needle Guidance GUI
1
Alignment indicator (plan view)
1
2
3
4
6
5
7
8
Summary of Contents for Anesus ME7T
Page 2: ......
Page 58: ...This page intentionally left blank ...
Page 154: ...This page intentionally left blank ...
Page 164: ...This page intentionally left blank ...
Page 182: ...This page intentionally left blank ...
Page 190: ...This page intentionally left blank ...
Page 208: ...This page intentionally left blank ...
Page 254: ...This page intentionally left blank ...
Page 264: ...This page intentionally left blank ...
Page 280: ...This page intentionally left blank ...
Page 311: ......
Page 312: ...P N 046 018839 00 5 0 ...