13 - 2
Operator’s Manual
13 Measurement, Comments and Body Mark
Distance (iScape)
Full screen
Within ±5% (for linear, wide-
convex, and phased probes)
Full screen
Within ±10% (for micro-
convex probes)
Table 13-2 Time/Motion Measurements
Parameter
Value Range
Error
Distance
Full screen
Within ±3%
Time
Timeline Display
Within ±2%
Heart rate
Timeline Display
Within ±4%
Velocity (PW mode)
10-200 cm/s (for non-
transcranial application)
When angle
≤
60°,
≤
5%
10-300 cm/s (for transcranial
application)
Within ±20%
Velocity (CW mode)
10-200 cm/s (for non-
transcranial application)
When angle
≤
60°,
≤
5%
10-300 cm/s (for transcranial
application)
Within ±20%
Table 13-3
Function
Measurement item
Error
Smart IVC
Dmax
Within ±10%
Dmin
Within ±10%
CI
Within ±10%
DI
Within ±10%
IVCV
Within ±10%
Smart VTI
LVOT Vmax
Within ±10%
LVOT VTI
Within ±10%
LVOT HR
Within ±10%
LVOT SV
Within ±10%
LVOT CO
Within ±10%
Smart B-line
Percent
Within ±10%
Dist
Within ±10%
Avg Dist
Within ±10%
B lines
0
Table 13-1 Error of 2D Images
Parameter
Value Range
Error
Summary of Contents for Anesus ME7T
Page 2: ......
Page 58: ...This page intentionally left blank ...
Page 154: ...This page intentionally left blank ...
Page 164: ...This page intentionally left blank ...
Page 182: ...This page intentionally left blank ...
Page 190: ...This page intentionally left blank ...
Page 208: ...This page intentionally left blank ...
Page 254: ...This page intentionally left blank ...
Page 264: ...This page intentionally left blank ...
Page 280: ...This page intentionally left blank ...
Page 311: ......
Page 312: ...P N 046 018839 00 5 0 ...