6 Image Acquisition
Operator’s Manual
6 - 9
6.6
Anatomical M Mode
For an image in the traditional M mode, the M-mark line goes along the beams transmitted from the
probe. Thus it is difficult to obtain a good plane for difficult-to-image patients who cannot be
moved easily. However, in the Anatomical M mode, you can manipulate the M-mark line and move
it to any position at desired angles. The system supports anatomical M scanning (including Free
Xros M mode and Free Xros CM mode) in 2D imaging modes (B, Color, Power and TVI mode).
CAUTION
Anatomical M Mode and Color Anatomical M mode images are provided for
reference only, not for confirming diagnoses. Compare the image with those of
other machines, or make diagnoses using non-ultrasound methods.
TIP:
Free Xros M imaging and Free Xros CM imaging are options.
6.6.1 Linear Anatomical M (Free Xros M)
Free Xros M imaging is supported on frozen B image, B+M image and B+Power/Color/TVI image.
Perform the following procedure:
1.
Adjust the probe and image to obtain the desired plane in real-time B mode or M mode.
Or select the B mode cine file to be observed.
2.
Tap [Free Xros M] on the touch screen to enter Free Xros M mode or press the user-defined
key to enter Free Xros M mode.
There are 3 M-mark lines available, each with a symbol of “A”, “B” or “C” at one end as
identification.
3.
Adjust the sampling line (single line or couple of lines) to obtain optimized images and
necessary information.
–
Tap [Show A], [Show B] or [Show C] on the touch screen to adjust the sampling line. The
corresponding sampling line and the Free Xros M image appear on the screen. Then,
activate the sampling line.
–
Tap [Display Cur.] or [Display All] on the touch screen to select whether to display the
image of the current M-mark line or all.
You can choose to display the sampling line on the current image or all.
–
Press <Set> to switch among the sampling lines and press <Cursor> to show the cursor.
4.
Adjust the image parameters to obtain optimized images.
5.
Press <B> to return to real-time B mode.
6.6.2 Free Xros CM (Curved Anatomical M-Mode)
In Free Xros CM mode, the distance/time curve is generated from the sample line manually
depicted anywhere on the image. Free Xros CM is used for TVI and TEI modes.
Summary of Contents for Anesus ME7T
Page 2: ......
Page 58: ...This page intentionally left blank ...
Page 154: ...This page intentionally left blank ...
Page 164: ...This page intentionally left blank ...
Page 182: ...This page intentionally left blank ...
Page 190: ...This page intentionally left blank ...
Page 208: ...This page intentionally left blank ...
Page 254: ...This page intentionally left blank ...
Page 264: ...This page intentionally left blank ...
Page 280: ...This page intentionally left blank ...
Page 311: ......
Page 312: ...P N 046 018839 00 5 0 ...