1 Safety Precautions
Operator’s Manual
1 - 7
•
When using a probe, pay attention to the status of the ultrasound image.
Do not use the probe to perform image acquisition when the image is
frozen.
CAUTION
•
When using the probe, wear sterile gloves to prevent infection.
•
Be sure to use sterile ultrasound gel. Please use the ultrasound gel com-
pliant with the relevant local regulations. And manage the ultrasound gel
properly to ensure that it does not become a source of infection.
•
In normal diagnostic ultrasound mode, there is no danger of a normal-
temperature burn; however, keeping the probe on the same region of the
patient for a long time may cause such a burn.
•
Do not use the carrying case for storing the transducer. If the carrying case
is used for storage, it may become a source of infection.
•
It is required to practice ALARA when operating ultrasound system.
Minimize the acoustic power without compromising the quality of images.
•
The probe and accessories supplied with it are not delivered disinfected or
sterilized. Sterilization (or high-level disinfect) before use is required.
•
Disposable components should be packaged sterile and for single-use
only. Do not use if integrity of packaging violated or if expiration date has
passed. Please use the disposable components compliant with the relevant
local regulations.
•
Please use the disinfection or sterilization solution recommended in this
operator's manual; otherwise Mindray will not be liable for damage caused
by other solutions. If you have any questions, please contact Mindray
Customer Service Department or sales representative.
•
Do not use pre-lubricated condoms as a sheath. Lubricant may not be
compatible with the probe material and damage may result.
•
The damage of the transducer may be caused by the contact of improper
gel or cleaner:
– DO NOT dip the transducer in the strong polar solution of ethanol,
chloride of lime, ammonium chloride, acetone and formaldehyde.
– DO NOT contact the transducer with solution or ultrasound gel
containing oily medium such as mineral oil or lanoline.
NOTE:
•
Read the following precautions to prevent the probe from malfunction:
–
Before connecting or disconnecting the probe, freeze or turn off the diagnostic ultrasound
system.
–
Clean and disinfect the probe before and after each examination.
–
After the examination, wipe off the ultrasound gel thoroughly. Otherwise, the ultrasound
gel may solidify and the image quality would be degraded.
Summary of Contents for Anesus ME7T
Page 2: ......
Page 58: ...This page intentionally left blank ...
Page 154: ...This page intentionally left blank ...
Page 164: ...This page intentionally left blank ...
Page 182: ...This page intentionally left blank ...
Page 190: ...This page intentionally left blank ...
Page 208: ...This page intentionally left blank ...
Page 254: ...This page intentionally left blank ...
Page 264: ...This page intentionally left blank ...
Page 280: ...This page intentionally left blank ...
Page 311: ......
Page 312: ...P N 046 018839 00 5 0 ...