16 - 26
Operator’s Manual
16 Probes and Biopsy
a.
Pull the lock pin and open up the V-shaped cover to expose the needle.
b.
Separate the bracket and the probe from the needle.
5.
Remove the needle-guided bracket:
a.
Screw the pinch nut to release the needle-guided bracket.
b.
Separate the bracket and the probe.
NGB-025 Metal-needle undetachable needle-guided bracket
Perform the following procedure:
1.
Install the needle-guided bracket:
a.
Put on the sterile probe sheath.
1
Pinch nut
2
Intra-cavity probe
3
Lower clamp
4
Location hole
5
Front slot
6
Front clamp
7
Location clamp
8
Upper clamp
1
2
3
4
5
6
7
8