16 - 42
Operator’s Manual
16 Probes and Biopsy
Controls are as shown in Table 16-6.
eS.Navi Settings Menu
Tap [eSpacial Navi] > [eS.Navi Settings] to enter the “eS.Navi Settings” menu, and tap the items to
be displayed on the screen.
16.4.3 Preparation Before Needle Guidance
NOTE:
Before and during calibration, keep the probe away from the magnetic sources, such as metal
objects, electrical motors, switching power supplies, nerve stimulators and similar medical devices.
Perform the following procedure:
1.
Scan the target tissue and adjust the image parameters.
2.
Check the field strength to ensure that the filed strength indicator is
. If the filed strength is
low, hold the probe and tap [Calibrate] on “eS.Navi Needle” menu. Do not move the probe
during the calibration.
3.
Magnetize a needle cap.
a.
Place an appropriate sterile needle cap (length: 3.5cm to 4.5cm, diameter < 7 mm, with
bottom closed) into the magnetizer.
b.
Introduce a needle into the cap, and ensure that the needle tip contacts the bottom of the
needle cap.
Table 16-6 eSpacial Navi Needle settings
Item
Description
Needle
Select the desired needle from the drop-down list.
You can set the most frequently used needle to a favorite one.
Display Favourite Only
After it is ticked, only the favorite needles are displayed on the needles
list.
Auto Optimize
After it is ticked, the system adjusts focus position automatically
according to the position of the needle tip, and enable or disable
ExFOV function.
Indicator Colors
Description
Green
Low magnetic interference: appropriate for needle guidance.
Yellow
Medium magnetic interference: be cautious for needle guidance.
Red
High magnetic interference: do not perform needle guidance.
Summary of Contents for Anesus ME7T
Page 2: ......
Page 58: ...This page intentionally left blank ...
Page 154: ...This page intentionally left blank ...
Page 164: ...This page intentionally left blank ...
Page 182: ...This page intentionally left blank ...
Page 190: ...This page intentionally left blank ...
Page 208: ...This page intentionally left blank ...
Page 254: ...This page intentionally left blank ...
Page 264: ...This page intentionally left blank ...
Page 280: ...This page intentionally left blank ...
Page 311: ......
Page 312: ...P N 046 018839 00 5 0 ...