4 - 34
Operator’s Manual
4 Setup
HL7 service setting for Worklist is described as follows:
4.8
Network Preset
4.8.1 Network Settings
Transmission Encryption
After accessing the network, click [VPN Config] to enter the “VPN Config” interface.
Item
Description
Device Name
After you set the server (s) in DICOM Server Setting screen, the name
(s) will appear in the drop-down list, select the name of the Worklist
server.
Service Name
Default is server-HL7Query, and it can be modified.
AE Title
Application Entity title.here, it should be consistent with that of the
HL7 server.
Port
DICOM communication port, 104 by default.Here, the port should be
consistent with that of the HL7 server port.
Maximum Retries
Reserved feature.
Interval Time(s)
Reserved feature.
Timeout
Refers to the amount of time after which the system will stop trying to
establish a connection to the service.
Listen mode
This function enables the ultrasound system to use the listen port for
data receiving.
Listen port
Port for ultrasound system to receive data after the listen mode function
is activated. Here, the port should be consistent with that of the HL7
server port.
For details of listen port setting, refer to settings in the server.
Summary of Contents for Anesus ME7T
Page 2: ......
Page 58: ...This page intentionally left blank ...
Page 154: ...This page intentionally left blank ...
Page 164: ...This page intentionally left blank ...
Page 182: ...This page intentionally left blank ...
Page 190: ...This page intentionally left blank ...
Page 208: ...This page intentionally left blank ...
Page 254: ...This page intentionally left blank ...
Page 264: ...This page intentionally left blank ...
Page 280: ...This page intentionally left blank ...
Page 311: ......
Page 312: ...P N 046 018839 00 5 0 ...