NGB-043 Metal-needle detachable needle-guided bracket
Perform the following procedure:
1.
Install the needle-guided bracket:
a.
Put on the sterile probe sheath.
b.
Hold the probe in one hand, select the correct needle-guided bracket and hold it with the
other hand. Match the grooves of the needle-guided bracket to the tabs of the probe.
c.
Mount the bracket onto the probe, and rotate the locking nut to fix the bracket.
2.
Insert the biopsy needle:
a.
Insert a biopsy needle into the guiding block hole.
b.
Adjust the biopsy angle of the needle, and ensure that the needle can be adjusted
continuously within the range of 30º to 85º.
3.
Release the needle from the bracket:
Hold the probe, and remove the biopsy needle form the guiding hole of the guiding block.
4.
Remove the needle-guided bracket:
1
Locking nut of left and right clamps
2
Groove
3
Guide frame
4
Guide block hole
5
Clamps
1
2
3
4
5