Operator’s Manual
C - 1
C
Electrical Safety Inspection
The following electrical safety tests are recommended as part of a comprehensive preventive
maintenance program. They are a proven means of detecting abnormalities that, if undetected,
could prove dangerous to either the patient or the operator. Additional tests may be required
according to local regulations.
All tests can be performed using commercially-available safety analyzer test equipment. These
procedures assume the use of a 601PROXL International Safety Analyzer or equivalent safety
analyzer. Other popular testers which comply with IEC 60601-1 and are used in Europe, such as
Fluke, Metron or Gerb, may require modifications to the procedure. Follow the analyzer
manufacturer's instructions.
An electrical safety inspection should be periodically performed every two years. The safety
analyzer is also an excellent troubleshooting tool for detecting abnormalities in line voltage and
grounding, as well as total current loads.
NOTE:
Make sure the safety analyzer is authorized and complies with the requirements of IEC 61010-1.
Follow the analyzer manufacturer's instructions.
C.1
Power Cord Plug
C.1.1 The Power Plug
Test Item
Acceptance Criteria
The power plug
The power plug pins
No broken or bent pins. No discolored pins.
The plug body
No physical damage to the plug body.
The strain relief
No physical damage to the strain relief. No plug
warmth when device is in use.
The power plug
No loose connections.
The power cord
No physical damage to the cord. No deterioration to
the cord.
For devices with detachable power cords, inspect the
connection with the device.
For devices with non-detachable power cords, inspect
the strain relief at the device.
Summary of Contents for Anesus ME7T
Page 2: ......
Page 58: ...This page intentionally left blank ...
Page 154: ...This page intentionally left blank ...
Page 164: ...This page intentionally left blank ...
Page 182: ...This page intentionally left blank ...
Page 190: ...This page intentionally left blank ...
Page 208: ...This page intentionally left blank ...
Page 254: ...This page intentionally left blank ...
Page 264: ...This page intentionally left blank ...
Page 280: ...This page intentionally left blank ...
Page 311: ......
Page 312: ...P N 046 018839 00 5 0 ...