I EMC Guidance and Manufacturer’s Declaration
Operator’s Manual
I - 3
RF emissions
CISPR 11
Class B
The system is suitable for use in all establishments
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage Fluctuations/
Flicker Emissions
IEC 61000-3-3
Compliance
Table I-1
GUIDANCE AND MINDRAY DECLARATION-ELECTROMAGNETIC EMISSIONS
The system is intended for use in the electromagnetic environment specified below. The
customer or the user of system should assure that it is used in such an environment.
EMISSIONS TEST
COMPLIANCE
ELECTROMAGNETIC ENVIROMENT - GUIDANCE
Table I-2
GUIDANCE AND MINDRAY DECLARATION-ELECTROMAGNETIC IMMUNITY
The system is intended for use in the electromagnetic environment specified below. The
customer or the user of system should assure that it is used in such an environment.
IMMUNITY TEST
IEC 60601
TEST LEVEL
COMPLIANCE LEVEL ELECTROMAGNETIC
ENVIRONMENT-
GUIDANCE
Electrostatic
Discharge(ESD)
IEC 61000-4-2
±8 kV contact;
±15 kV air
±8 kV contact;
±15 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Electrical fast Transient /
burst
IEC 61000-4-4
±2 kV for power supply
lines;
±1 kV for input/output
lines
±2 kV for power supply
lines;
±1 kV for input/output
lines
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV line(s) to line(s);
±2 kV line(s) to earth
±1 kV line(s) to line(s);
±2 kV line(s) to earth
Mains power quality
should be that of a typical
commercial or hospital
environment.
Voltage dips, Short
interruptions and voltage
variation on power supply
input voltage
IEC 61000-4-11
0 % U
T
; 0,5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and 315°
0 % U
T
; 1 cycle
70% U
T
for 25/30 cycle
at 0°
0 % U
T
; 250/300 cycle
0 % U
T
; 0,5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and 315°
0 % U
T
; 1 cycle
70% U
T
for 25/30 cycle
at 0°
0 % U
T
; 250/300 cycle
Mains power quality
should be that of a typical
commercial or hospital
environment. If you
require continued
operation during power
mains interruptions, it is
recommended that our
product be powered from
an uninterruptible power
supply or a battery.
Summary of Contents for Anesus ME7T
Page 2: ......
Page 58: ...This page intentionally left blank ...
Page 154: ...This page intentionally left blank ...
Page 164: ...This page intentionally left blank ...
Page 182: ...This page intentionally left blank ...
Page 190: ...This page intentionally left blank ...
Page 208: ...This page intentionally left blank ...
Page 254: ...This page intentionally left blank ...
Page 264: ...This page intentionally left blank ...
Page 280: ...This page intentionally left blank ...
Page 311: ......
Page 312: ...P N 046 018839 00 5 0 ...