Proline Prowirl F 200 PROFIBUS PA
Commissioning
Hauser
93
Parameter
Prerequisite
Description
Selection / User
entry
Factory setting
Mol% CO2
The following conditions are
met:
In the
Select medium
parameter, the
Gas
option is
selected.
– In the
Select gas type
parameter, the
Gas mixture
option is selected and in the
Gas mixture
parameter, the
Carbon dioxide CO2
option
is selected.
Or
– In the
Select gas type
parameter, the
Natural gas
option is selected.
Enter amount of substance for
the gas mixture.
0 to 100 %
0 %
Mol% H2
The following conditions are
met:
In the
Select medium
parameter, the
Gas
option is
selected.
– In the
Select gas type
parameter, the
Gas mixture
option is selected and in the
Gas mixture
parameter, the
Hydrogen H2
option is
selected.
Or
– In the
Select gas type
parameter, the
Natural gas
option is selected and in the
Density calculation
parameter, the
AGA Nx19
option is
not
selected.
Enter amount of substance for
the gas mixture.
0 to 100 %
0 %
Mol% H2O
The following conditions are
met:
• In the
Select medium
parameter, the
Gas
option is
selected.
• In the
Select gas type
parameter, the
Natural gas
option is selected.
• In the
Density calculation
parameter, the
ISO 12213-
2
option is selected.
Enter amount of substance for
the gas mixture.
0 to 100 %
0 %
Mol% H2S
The following conditions are
met:
In the
Select medium
parameter, the
Gas
option is
selected.
– In the
Select gas type
parameter, the
Gas mixture
option is selected and in the
Gas mixture
parameter, the
Hydrogen sulfide H2S
option is selected.
Or
– In the
Select gas type
parameter, the
Natural gas
option is selected and in the
Density calculation
parameter, the
ISO 12213-
2
option is selected.
Enter amount of substance for
the gas mixture.
0 to 100 %
0 %