Vibration resistance
Vibration, sinusoidal according to IEC 60068-2-6
• Order code for "Housing", option C "GT20 two-chamber, aluminum, coated, compact", J
"GT20 two-chamber, aluminum, coated, remote", K "GT18 two-chamber, 316L, remote"
– 2 to 8.4 Hz, 7.5 mm peak
– 8.4 to 500 Hz, 2 g peak
• Order code for "Housing", option B "GT18 two-chamber, 316L, compact"
– 2 to 8.4 Hz, 3.5 mm peak
– 8.4 to 500 Hz, 1 g peak
Vibration broad-band random, according to IEC 60068-2-64
• Order code for "Housing", option C "GT20 two-chamber, aluminum, coated, compact", J
"GT20 two-chamber, aluminum, coated, remote", K "GT18 two-chamber, 316L, remote"
– 10 to 200 Hz, 0.01 g
2
/Hz
– 200 to 500 Hz, 0.003 g
2
/Hz
– Total 2.7 g rms
• Order code for "Housing", option B "GT18 two-chamber, 316L, compact"
– 10 to 200 Hz, 0.003 g
2
/Hz
– 200 to 500 Hz, 0.001 g
2
/Hz
– Total 1.54 g rms
Shock resistance
Shock, half-sine according to IEC 60068-2-27
• Order code for "Housing", option C "GT20 two-chamber, aluminum, coated, compact", J
"GT20 two-chamber, aluminum, coated, remote", K "GT18 two-chamber, 316L, remote"
6 ms, 50 g
• Order code for "Housing", option B "GT18 two-chamber, 316L, compact"
6 ms, 30 g
Impact resistance
Rough handling shocks according to IEC 60068-2-31
Electromagnetic
compatibility (EMC)
Details are provided in the Declaration of Conformity.
16.9 Process
Medium temperature range
DSC sensor
Order code for "Sensor version; DSC sensor; measuring tube"
Option Description
Medium temperature range
AA
Volume; 316L; 316L
–40 to +260 °C (–40 to +500 °F), stainless steel
AB
Volume; Alloy C22; 316L
AC
Volume; Alloy C22; Alloy C22
–40 to +260 °C (–40 to +500 °F), stainless steel
BA
Volume high-temperature; 316L; 316L
–200 to +400 °C (–328 to +752 °F), stainless steel
BB
Volume high-temperature; Alloy C22; 316L
CA
Mass; 316L; 316L
–200 to +400 °C (–328 to +752 °F), stainless steel
CB
Mass; Alloy C22; 316L
CC
Mass; Alloy C22; Alloy C22
–40 to +260 °C (–40 to +500 °F), stainless steel
1)
Capacitance sensor