Proline Prowirl F 200 PROFIBUS PA
Commissioning
Hauser
85
Dynamic viscosity
‣
Gas composition
Parameter overview with brief description
Parameter
Prerequisite
Description
Selection / User
entry
Factory setting
Enthalpy type
The following conditions are
met:
• In the
Select gas type
parameter, the
User-
specific gas
option is
selected.
Or
• In the
Select liquid type
parameter, the
User-
specific liquid
option is
selected.
Define which kind of enthalpy
is used.
• Heat
• Calorific value
Heat
Calorific value type
The
Calorific value type
parameter is visible.
Select calculation based on
gross calorific value or net
calorific value.
• Gross calorific
value volume
• Net calorific value
volume
• Gross calorific
value mass
• Net calorific value
mass
Gross calorific value
mass
Reference combustion temperature
The
Reference combustion
temperature
parameter is
visible.
Enter reference combustion
temperature to calculate the
natural gas energy value.
Dependency
The unit is taken from the
Temperature unit
parameter
–200 to 450 °C
20 °C
Reference density
The following conditions are
met:
• In the
Select gas type
parameter, the
User-
specific gas
option is
selected.
Or
• In the
Select liquid type
parameter, the
Water
option or
User-specific
liquid
option is selected.
Enter fixed value for reference
density.
Dependency
The unit is taken from the
Density unit
parameter
0.01 to 15
000 kg/m³ 1
000 kg/m³
Reference gross calorific value
The following conditions are
met:
• In the
Select medium
parameter, the
Gas
option is
selected.
• In the
Select gas type
parameter, the
Natural gas
option is selected.
• In the
Density calculation
parameter, the
ISO 12213-
3
option is selected.
Enter reference gross calorific
value of the natural gas.
Dependency
The unit is taken from the
Calorific value unit
parameter
Positive floating-
point number
50
000 kJ/Nm³