Commissioning
Proline Prowirl F 200 PROFIBUS PA
120
Hauser
3. Pull out the display module with a gentle rotational movement. To make it easier to
access the write protection switch, attach the display module to the edge of the
electronics compartment.
Display module is attached to the edge of the electronics compartment.
A0032236
4. Setting the write protection switch (WP) on the main electronics module to the
ON
position enables hardware write protection. Setting the write protection switch (WP)
on the main electronics module to the
OFF
position (factory setting) disables
hardware write protection.
If the hardware write protection is enabled: The
Hardware locked
option is
displayed in the
Locking status
parameter . In addition, on the local display the
-symbol appears in front of the parameters in the header of the operational
display and in the navigation view.
X X X X
X X X
X
X
20.50
XX
XX
A0029425
If the hardware write protection is disabled: No option is displayed in the
Locking
status
parameter . On the local display, the -symbol disappears from in front
of the parameters in the header of the operational display and in the navigation
view.
5. Feed the cable into the gap between the housing and main electronics module and
plug the display module into the electronics compartment in the desired direction
until it engages.
6. Reverse the removal procedure to reassemble the transmitter.
10.8 Application-specific commissioning
10.8.1 Steam application
Select medium
Navigation:
Setup → Medium selection
1. Call up the
Medium selection
wizard.
Summary of Contents for Proline Prowirl F 200
Page 224: ...www addresses endress com...