Proline Prowirl F 200 PROFIBUS PA
Commissioning
Hauser
103
Example
Influence of the diameter mismatch without using the correction function:
• Mating pipe DN 100 (4"), schedule 80
• Device flange DN 100 (4"), schedule 40
• This installation position results in a diameter mismatch of 5 mm (0.2 in). If the
correction function is not used, an additional measuring uncertainty of approx. 2 % o.r.
must be expected.
• If the basic conditions are met and the feature is enabled, the additional measuring
uncertainty is 1 % o.r.
10.5.4 Configuring the pulse/frequency/switch output
The
Pulse/frequency/switch output
wizard guides you systematically through all the
parameters that can be set for configuring the selected output type.
Navigation
"Setup" menu → Pulse/frequency/switch output
Pulse/frequency/switch output
Operating mode
Parameter overview with brief description
Parameter
Description
Selection
Factory setting
Operating mode
Define the output as a pulse, frequency or
switch output.
• Pulse
• Frequency
• Switch
Pulse
Configuring the pulse output
Navigation
"Setup" menu → Pulse/frequency/switch output
Pulse/frequency/switch output
Assign pulse output 1
Value per pulse
Pulse width
Failure mode
Invert output signal