Diagnostics and troubleshooting
Proline Prowirl F 200 PROFIBUS PA
160
Hauser
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
412 Processing Download
Download active, please wait
• Calculated saturated
steam pressure
• Density
• Energy flow
• Flow velocity
• Heat flow difference
• Low flow cut off
• Mass flow
• Total mass flow
• Switch output status
• Pressure
• Reynolds number
• Specific volume
• Corrected volume flow
• Steam quality
• Degrees of superheat
• Temperature
• Volume flow
Measured variable status
Quality
Uncertain
Quality substatus
Initial value
Coding (hex)
0x4C to 0x4F
Status signal
C
Diagnostic behavior
Warning
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
437 Configuration incompatible
1. Restart device
2. Contact service
• Calculated saturated
steam pressure
• Density
• Energy flow
• Flow velocity
• Heat flow difference
• Low flow cut off
• Mass flow
• Total mass flow
• Switch output status
• Pressure
• Reynolds number
• Specific volume
• Corrected volume flow
• Steam quality
• Degrees of superheat
• Temperature
• Volume flow
Measured variable status
Quality
Bad
Quality substatus
Maintenance alarm
Coding (hex)
0x24 to 0x27
Status signal
F
Diagnostic behavior
Alarm
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
438 Dataset
1. Check data set file
2. Check device configuration
3. Up- and download new configuration
• Calculated saturated
steam pressure
• Density
• Energy flow
• Flow velocity
• Heat flow difference
• Low flow cut off
• Mass flow
• Total mass flow
• Switch output status
• Pressure
• Reynolds number
• Specific volume
• Corrected volume flow
• Steam quality
• Degrees of superheat
• Temperature
• Volume flow
Measured variable status
Quality
Uncertain
Quality substatus
Maintenance demanded
Coding (hex)
0x68 to 0x6B
Status signal
M
Diagnostic behavior
Warning