Proline Prowirl F 200 PROFIBUS PA
Commissioning
Hauser
99
Parameter overview with brief description
Parameter
Prerequisite
Description
Selection / User
entry
Factory setting
External value
With order code for "Sensor
version":
Option "Mass (integrated
temperature measurement)"
Assign variable from external
device to process variable.
Selection
NOTE!
If pressure is the selected
option, the pressure is read
in externally by means of a
pressure transmitter.
The pressure must be read in
the unit Pascal so that pressure
compensation can be read in
correctly.
▸ Select the
Pa
option in the
Pressure unit
parameter.
For detailed information
on the calculation of the
measured variables with
steam: → 125
For detailed information
on setting the parameter
in steam applications,
see the Special
Documentation for the
Wet Steam Detection
and
Wet Steam
Measurement
application package
• Off
• Pressure
• Gauge pressure
• Density
• Temperature
• 2nd temperature
delta heat
Off
Atmospheric pressure
In the
External value
parameter, the
Gauge
pressure
option is selected.
Enter atmospheric pressure
value to be used for pressure
correction.
Dependency
The unit is taken from the
Pressure unit
parameter
0 to 250 bar
1.01325 bar
Delta heat calculation
The
Delta heat calculation
parameter is visible.
Calculates the transferred heat
of a heat exchanger (= delta
heat).
• Off
• Device on cold side
• Device on warm
side
Device on warm side
Fixed density
With order code for "Sensor
version":
• Option "Volume"
or
• Option "Volume high
temperature"
Enter fixed value for medium
density.
Dependency
The unit is taken from the
Density unit
parameter.
0.01 to 15
000 kg/m³ 1
000 kg/m³
Fixed temperature
–
Enter a fixed value for process
temperature.
Dependency
The unit is taken from the
Temperature unit
parameter
–200 to 450 °C
20 °C
2nd temperature delta heat
The
2nd temperature delta
heat
parameter is visible.
Enter 2nd temperature value
to calculate the delta heat.
Dependency
The unit is taken from the
Temperature unit
parameter
–200 to 450 °C
20 °C
Summary of Contents for Proline Prowirl F 200
Page 224: ...www addresses endress com...