Proline Prowirl F 200 PROFIBUS PA
Commissioning
Hauser
125
flow. The error resulting from this assumption (typically 1 to 5 %) is often considerably
smaller than the error caused by inaccurate composition data. This method should not be
used for condensing gases (e. g. saturated steam).
Select medium
Navigation:
Setup → Medium selection
1. Call up the
Medium selection
wizard.
2. In the
Select medium
parameter, select the
Gas
option.
3. In the
Select gas type
parameter, select the
User-specific gas
option.
4. For non-flammable gas:
In the
Enthalpy type
parameter, select the
Heat
option.
Configuring fluid properties
Navigation:
Setup → Advanced setup → Medium properties
5. Call up the
Medium properties
submenu.
6. In the
Reference density
parameter, enter the reference density of the fluid.
7. In the
Reference pressure
parameter, enter the reference pressure of the fluid.
8. In the
Reference temperature
parameter, enter the fluid temperature associated
with the reference density.
9. In the
Reference Z-factor
parameter, enter the value
1
.
10. If specific heat capacity is to be measured:
In the
Specific heat capacity
parameter, enter the heat capacity of the fluid.
11. In the
Z-factor
parameter, enter the value
1
.
12. In the
Dynamic viscosity
parameter, enter the viscosity of the fluid under operating
conditions.
10.8.4 Calculation of the measured variables
A flow computer can be found in the electronics of the measuring device with order code
for "Sensor version", option "mass (integrated temperature measurement)" and option "mass
(integrated pressure/temperature measurement)". This computer can calculate the
following secondary measured variables directly from the primary measured variables
recorded using the pressure value (entered or external) and/or temperature value
(measured or entered).
Mass flow and corrected volume flow
Medium Fluid
Standards
Explanation
Steam
Water vapor
IAPWS-IF97/
ASME
• For integrated temperature measurement
• For fixed process pressure, pressure measured directly at the
meter body or if the pressure is read in via PROFIBUS PA
Gas
Single gas
NEL40
For fixed process pressure, pressure measured directly at the meter
body or if the pressure is read in via PROFIBUS PA
Gas mixture
NEL40
Air
NEL40
Natural gas
ISO 12213-2
• Contains AGA8-DC92
• For fixed process pressure, pressure measured directly at the
meter body or if the pressure is read in via PROFIBUS PA