Diagnostics and troubleshooting
Proline Prowirl F 200 PROFIBUS PA
148
Hauser
The diagnostic information is grouped as follows:
• Diagnostic information pertaining to the sensor: diagnostic number 000 to 199
• Diagnostic information pertaining to the electronics: diagnostic number 200 to 399
• Diagnostic information pertaining to the configuration: diagnostic number 400 to 599
• Diagnostic information pertaining to the process: diagnostic number 800 to 999
Depending on the group in which the diagnostic information is located, the following
measured value status and device status are firmly assigned to the particular diagnostic
behavior:
Diagnostic information pertaining to the sensor: diagnostic number 000 to 199
Diagnostic behavior
(configurable)
Measured value status (fixed assignment)
Device diagnosis
(fixed assignment)
Quality
Quality
Substatus
Coding
(hex)
Category
(NE107)
Alarm
BAD
Maintenance
alarm
0x24 to 0x27
F
(Failure)
Maintenance
alarm
Warning
GOOD
Maintenance
demanded
0xA8 to 0xAB
M
(Maintenance)
Maintenance
demanded
Logbook entry only
GOOD
ok
0x80 to 0x8E
–
–
Off
Diagnostic information pertaining to the electronics: diagnostic number 200 to 399
Diagnostic behavior
(configurable)
Measured value status (fixed assignment)
Device diagnosis
(fixed assignment)
Quality
Quality
Substatus
Coding
(hex)
Category
(NE107)
Alarm
BAD
Maintenance
alarm
0x24 to 0x27
F
(Failure)
Maintenance
alarm
Warning
Logbook entry only
GOOD
ok
0x80 to 0x8E
–
–
Off
Diagnostic information pertaining to the configuration: diagnostic number 400 to 599
Diagnostic behavior
(configurable)
Measured value status (fixed assignment)
Device diagnosis
(fixed assignment)
Quality
Quality
Substatus
Coding
(hex)
Category
(NE107)
Alarm
BAD
Process
related
0x28 to 0x2B
F
(Failure)
Invalid process
condition
Warning
UNCERTA
IN
Process
related
0x78 to 0x7B
S
(Out of
specification)
Invalid process
condition
Logbook entry only
GOOD
ok
0x80 to 0x8E
–
–
Off
Summary of Contents for Proline Prowirl F 200
Page 224: ...www addresses endress com...